Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type
Project begin
2011-07-13
WI
00205243
Planned document number
DIN EN ISO 81060-2
Abstract
This part of ISO 81060 specifies the requirements and methods for the clinical validation of me equipment used used for the intermittent non-invasive automatic estimation of the arterial blood pressure by utilizing a cuff. This part of ISO 81060 is applicable to all sphygmomanometers that sense or display pulsations, flow or sounds for the estimation, display or recording of blood pressure. These sphygmomanometers need not have automatic cuff inflation. This part of ISO 81060 covers sphygmomanometers intended for use in all patient populations (e.g. all age and weight ranges), and all conditions of use (e.g. ambulatory blood pressure monitoring, stress testing blood pressure monitoring and blood pressure monitors for the home healthcare environment of self-measurement).
Responsible Committee
NA 027-02-14 AA Non-invasive sphygmomanometer